Comparison of Initial Mechanical Properties of 4 Hamstring Graft Femoral Fixation Systems Using Nonpermanent Hardware for Anterior Cruciate Ligament Reconstruction: An In Vitro Animal Study

dc.centroEscuela de Ingenierías Industrialeses_ES
dc.contributor.authorEspejo-Baena, Alejandro
dc.contributor.authorEzquerro-Juanco, Francisco
dc.contributor.authorPérez-de-la-Blanca-Cobos, Ana María
dc.contributor.authorFernández-Serrano, José Luis
dc.contributor.authorNadal-Martínez, Fernando
dc.contributor.authorMontañez Heredia, Elvira
dc.date.accessioned2025-01-30T19:57:35Z
dc.date.available2025-01-30T19:57:35Z
dc.date.issued2006
dc.departamentoIngeniería Mecánica, Térmica y de Fluidos
dc.description.abstractPurpose: To compare the initial mechanical characteristics of 4 systems used to fix tendons to the femur during anterior cruciate ligament reconstruction. Methods: A total of 32 porcine femurs were used to study the following fixation systems: Bioabsorbable interference screw (Stryker, Kalamazoo, MI), Bio-Transfix Cross-pin (Arthrex, Naples, FL), Biosteon Cross-pin (Stryker), and a fixation technique based on wrapping the graft around the femoral condyle itself, thus allowing it to be fixed in place without the use of any hardware. The mechanical characteristics of each system were obtained by a preconditioned failure tensile test. Results: The yield load values (990.9 ± 242.6 N for Bio-Transfix, 905.1 ± 158.8 N for Biosteon Cross-pin, 684.4 119.7 N for the without-hardware system (WHS), and 369.4 ± 120.1 N for the interference screw) revealed significant differences between the techniques that used cross-pins and the other 2 techniques (P < .006) on the one hand, and between the without hardware technique and the interference screw (P < .004) on the other. The stiffness of the 2 cross-pin fixation systems (117.6 ± 22.5 N for Bio-Transfix and 112.6 ± 22.5 N for Biosteon) was greater (P < .01) than those of the other systems (79.4 ± 15.2 N for the WHS and 68.5 ± 13 N for the interference screw). Conclusions: The initial biomechanical properties of the 2 cross-pin fixation systems proved to be superior to those of the other 2 systems studied. The WHS fixation system exhibited better mechanical properties than its interference screw counterpart. Clinical Relevance: The better initial mechanical characteristics encountered using the BioTransfix and Biosteon Cross-pin systems indicate that these systems are better equipped to bear the loads generated by aggressive rehabilitation. The WHS fixation system provides an alternative to interference screw fixation.es_ES
dc.identifier.citationAlejandro Espejo-Baena, Francisco Ezquerro, Ana Pérez de la Blanca, José Serrano-Fernandez, Fernando Nadal, Elvira Montañez-Heredia, Comparison of Initial Mechanical Properties of 4 Hamstring Graft Femoral Fixation Systems Using Nonpermanent Hardware for Anterior Cruciate Ligament Reconstruction: An In Vitro Animal Study, Arthroscopy: The Journal of Arthroscopic & Related Surgery, Volume 22, Issue 4, 2006, Pages 433-440, ISSN 0749-8063es_ES
dc.identifier.doi10.1016/j.arthro.2005.09.021
dc.identifier.urihttps://hdl.handle.net/10630/37457
dc.language.isoenges_ES
dc.publisherElsevieres_ES
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internacional*
dc.rights.accessRightsopen accesses_ES
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.subjectBiomecánicaes_ES
dc.subject.otherHamstring graft fixationes_ES
dc.subject.otherBiomechanicses_ES
dc.subject.otherACL reconstructiones_ES
dc.subject.otherKneees_ES
dc.titleComparison of Initial Mechanical Properties of 4 Hamstring Graft Femoral Fixation Systems Using Nonpermanent Hardware for Anterior Cruciate Ligament Reconstruction: An In Vitro Animal Studyes_ES
dc.typejournal articlees_ES
dc.type.hasVersionSMURes_ES
dspace.entity.typePublication
relation.isAuthorOfPublicationd9e856f5-be26-4bfd-9b4f-17a0a1310827
relation.isAuthorOfPublication3a0a9c46-bd91-4113-a087-1258a7309555
relation.isAuthorOfPublication79ded3d3-bc18-40eb-ab6f-5a3241af32a7
relation.isAuthorOfPublication107d1693-14af-44b8-950f-5bfe32bd45e6
relation.isAuthorOfPublication.latestForDiscoveryd9e856f5-be26-4bfd-9b4f-17a0a1310827

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Comparison of Initial Mechanical Properties of 4 Hamstring Graft Femoral Fixation Systems.pdf
Size:
190.36 KB
Format:
Adobe Portable Document Format
Description:

Collections