Performance of a virtual source model for stereotactic radiosurgery with a dynamic micro-multileaf collimator.

dc.centroFacultad de Cienciases_ES
dc.contributor.authorGonzález-Infantes, Wilfredo
dc.contributor.authorAnguiano, Marta
dc.contributor.authorLallena, Antonio
dc.date.accessioned2025-01-28T08:22:37Z
dc.date.available2025-01-28T08:22:37Z
dc.date.issued2017-11-17
dc.departamentoFísica Aplicada I
dc.descriptionhttps://publishingsupport.iopscience.iop.org/questions/what-is-iop-publishings-green-open-access-policy/es_ES
dc.description.abstractA virtual source model recently developed for external radiotherapy, which includes two photon sources and two electron sources, is tested for stereotactic radiosurgery with a dynamic micro-multileaf collimator. A 6 MV Elekta Precise linac that incorporates a 3Dline L’Arancio dynamic micro-multileaf collimator was considered. The capabilities of the virtual source model have been tested against full simulations performed with the Monte Carlo simulation code PENELOPE using both PENELOPE itself and the rapid Monte Carlo code DPM. The reference 10 cm × 10 cm radiation field as well as small fields of 1.2 cm × 1.2 cm, 2.9 cm × 2.9 cm and 5.8 cm×5.8 cm were considered. The dynamic micro-multileaf collimator effect can be included in the virtual source model by reducing the width of the photon primary source by a factor 3. In all cases the virtual source model simulations were in very good agreement with the complete ones including the linac and collimator geometries.es_ES
dc.identifier.citationW González et al 2017 Biomed. Phys. Eng. Express 3 065008es_ES
dc.identifier.doi10.1088/2057-1976/aa84ca
dc.identifier.urihttps://hdl.handle.net/10630/37133
dc.language.isoenges_ES
dc.publisherIOPes_ES
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internacional*
dc.rights.accessRightsopen accesses_ES
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.subjectÓpticaes_ES
dc.subjectEstereoencefalotomíaes_ES
dc.subject.othervirtual source modeles_ES
dc.subject.othermicro-multileaf collimatores_ES
dc.subject.otherstereotactic radiosurgeryes_ES
dc.titlePerformance of a virtual source model for stereotactic radiosurgery with a dynamic micro-multileaf collimator.es_ES
dc.typejournal articlees_ES
dc.type.hasVersionAMes_ES
dspace.entity.typePublication
relation.isAuthorOfPublication6bb188ce-1d12-4894-a2ce-6b247cdc3d64
relation.isAuthorOfPublication.latestForDiscovery6bb188ce-1d12-4894-a2ce-6b247cdc3d64

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
paper-20.pdf
Size:
379.93 KB
Format:
Adobe Portable Document Format
Description:

Collections