Three-Layer Control for Active Wrists in Robotized Laparoscopic Surgery.
Loading...
Identifiers
Publication date
Reading date
Collaborators
Advisors
Tutors
Editors
Journal Title
Journal ISSN
Volume Title
Publisher
IEEE/RSJ International Conference on Intelligent Robots and Systems
Share
Department/Institute
Abstract
This paper is focused on the motion control problem for a laparoscopic surgery robot assistant with an actuated wrist. These assistants may apply non-desired efforts to the patient abdomen. Therefore, this article proposes a control methodology based on three feedback levels, which have been defined as layers. These layers control different aspects of the endoscope movement. A low level assures the dynamic of the robot assistant is performed accordingly. The mid level emulates a passive wrist behavior to avoid any efforts over the abdomen. An external high level deals with the global movement planning. This architecture also makes easier to analyze the stability of the whole system. Finally, a real in-vitro experiment has been implemented with an industrial robot in order to contrast the validity of this article procedure.
Description
This work was supported in part by the Spanish National projects DPI2007-62257 and the Regional one P07-TEP-02897
Versión posprint
Bibliographic citation
Bauzano-Nuñez E; Muñoz-Martinez V.F.; García-Morales I., Estebanez-Campos B. Three-Layer Control for Active Wrists in Robotized Laparoscopic Surgery. Proceedings of the IEEE International Conference on Intelligent Robots and Systems (IROS) 2009, pp. 2653-2658. https://doi.org/10.1109/IROS.2009.5354614