<?xml version="1.0" encoding="UTF-8"?><?xml-stylesheet type="text/xsl" href="static/style.xsl"?><OAI-PMH xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/ http://www.openarchives.org/OAI/2.0/OAI-PMH.xsd"><responseDate>2026-05-30T06:31:50Z</responseDate><request verb="GetRecord" identifier="oai:riuma.uma.es:10630/36623" metadataPrefix="oai_dc">https://riuma.uma.es/rest/oai/request</request><GetRecord><record><header><identifier>oai:riuma.uma.es:10630/36623</identifier><datestamp>2026-02-03T11:13:45Z</datestamp><setSpec>com_10630_2254</setSpec><setSpec>col_10630_37953</setSpec></header><metadata><oai_dc:dc xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:doc="http://www.lyncode.com/xoai" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/ http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
   <dc:title>Carbon and oxygen minibeam radiation therapy: An experimental dosimetric evaluation</dc:title>
   <dc:creator>Martínez Rovira, Immaculada</dc:creator>
   <dc:creator>González-Infantes, Wilfredo</dc:creator>
   <dc:creator>Brons, S.</dc:creator>
   <dc:creator>Prezado, Yolanda</dc:creator>
   <dc:subject>Radiobiología</dc:subject>
   <dc:subject>Hadrones</dc:subject>
   <dc:subject>Hadron therapy</dc:subject>
   <dc:subject>Minibeam radiation therapy (MBRT)</dc:subject>
   <dc:subject>Ccarbon and oxygen beams</dc:subject>
   <dc:description>Abstract&#xd;
Purpose: To perform dosimetric characterization of a minibeam collimator in both carbon and&#xd;
oxygen ion beams to guide optimal set up geometry and irradiation for future radiobiological studies&#xd;
using these beams.&#xd;
Methods: Carbon and oxygen minibeams were generated using a prototype tungsten multislit&#xd;
collimator presenting line apertures 700 μm wide, which are spaced 3500 μm centre-to-centre&#xd;
distance apart. Several radiation beam spots generated the desired field size of 15×15 mm2 and&#xd;
production of a 50 mm long spread out Bragg peak (SOBP) centered at 80 mm-depth in water. Dose&#xd;
evaluations were performed with two different detectors: a PTW microDiamond R© single crystal&#xd;
diamond detector and radiochromic films (EBT3). Peak-to-valley dose ratio (PVDR) values, output&#xd;
factors (OF), penumbras and full width at half maximum (FWHM) were measured.&#xd;
Results: Measured lateral dose profiles exhibited spatial fractionation of dose at depth in a water&#xd;
phantom in the expected form of peaks and valleys for both carbon and oxygen radiation fields.&#xd;
The diamond detector and radiochromic film provided measurements of PVDR in good agreement.&#xd;
PVDR values at shallow depth were about 60 and decreased to about 10 at 80 mm-depth in water.&#xd;
OF in the center of the SOBP was about 0.4; this value is larger than the corresponding one in&#xd;
proton minibeam radiation therapy measured using comparable collimator due to a reduced lateral&#xd;
scattering for carbon and oxygen minibeams.&#xd;
Conclusions: Carbon and oxygen minibeams may be produced by a mechanical collimator.&#xd;
PVDR values and output factors measured in this first study of these minibeam radiation types&#xd;
indicate there is potential for their therapeutic use. Optimization of minibeam collimator design&#xd;
and the number and size of focal spots for irradiation are advocated to improve PDVR values and&#xd;
dose distributions for each specific applied use.</dc:description>
   <dc:date>2025-01-21T10:28:42Z</dc:date>
   <dc:date>2025-01-21T10:28:42Z</dc:date>
   <dc:date>2017-05-29</dc:date>
   <dc:type>journal article</dc:type>
   <dc:type>AM</dc:type>
   <dc:identifier>Martínez‐Rovira, I., González, W., Brons, S., &amp; Prezado, Y. (2017). Carbon and oxygen minibeam radiation therapy: An experimental dosimetric evaluation. Medical Physics (Lancaster), 44(8), 4223–4229. https://doi.org/10.1002/mp.12383</dc:identifier>
   <dc:identifier>https://hdl.handle.net/10630/36623</dc:identifier>
   <dc:identifier>10.1002/mp.12383</dc:identifier>
   <dc:language>eng</dc:language>
   <dc:rights>Attribution-NonCommercial-NoDerivatives 4.0 Internacional</dc:rights>
   <dc:rights>http://creativecommons.org/licenses/by-nc-nd/4.0/</dc:rights>
   <dc:rights>open access</dc:rights>
   <dc:format>application/pdf</dc:format>
   <dc:publisher>Wiley</dc:publisher>
</oai_dc:dc>
</metadata></record></GetRecord></OAI-PMH>